Buy ASTM F (R) GUIDE FOR ACCELERATED AGING OF STERILE BARRIER SYSTEMS FOR MEDICAL DEVICES from SAI. Time versus Temperature. – Relationship. – Affects on products. • Case Studies. • ASTM F – Changes to revision. • Real Time Aging. 1 This guide is under the jurisdiction of ASTM Committee F02 on Primary Last previous edition approved in as F – 07().

Author: Shakazshura Shaktikus
Country: Samoa
Language: English (Spanish)
Genre: Marketing
Published (Last): 19 February 2005
Pages: 381
PDF File Size: 17.48 Mb
ePub File Size: 10.60 Mb
ISBN: 750-7-97259-636-5
Downloads: 37187
Price: Free* [*Free Regsitration Required]
Uploader: Yozshujin

The resulting creation of an expiration date or shelf life is based on the use of a conservative estimate of the aging factor for example, Q 10 and is tentative until the results of real time aging studies are completed on the sterile barrier system.

More aggressive AAFs may be used with documented evidence to show a correlation between real time and accelerated aging. Define aging test time intervals including time zero.

Fed-Ex 2 nd day, Account number. The expiry date of a medical product is commonly based on its manufacture date. For more information visit www. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

ASTM-F – Accelerated Aging – Medical Package Testing

Referenced Documents purchase separately The documents listed below are referenced within the subject standard but are not provided as part of the standard. Evaluate the package performance after accelerated 1f980 relative to the initial package requirements.

Select the Q10 value Define the desired shelf life of the package marketing and product needs, etc.

Accelerated aging studies can provide an alternative means. ASTM F procedure for accelerated aging is comprised of the following: The primary accelerated aging standards pertaining to sterile barrier systems for medical devices are:.

  ASTROPOLITIK DOLMAN PDF

To ensure that accelerated aging studies do truly represent real time effects, real time aging studies must be conducted in parallel to accelerated studies. Atsm to Practice D for standard conditions that may be used to challenge the sterile barrier system to realistic extremes in temperature and humidity conditions. In parallel, age samples at real-life sstm conditions TRT. If you have questions about our online quoting process or other questions or need assistance, call RMTS at Try out our Accelerated Aging Calculator!

Age samples at TAA. Evaluate package, or package performance, atsm both, after real time aging relative to the initial package requirements.

Please fill out this form and return with your Product Submission. Standards for Accelerated Aging The primary accelerated aging standards pertaining to sterile barrier systems for medical devices are: However, due to market conditions in which products become obsolete in a short time, and the need to get new products to market in the shortest possible time, real time aging studies do not meet this objective.

To receive an Asttm Detailed Quote, please fill-in the following fields. UPS Overnight, Account number.

For example, a 12 month shelf life is simulated based on 13 months. Stability testing using accelerated aging protocols shall be regarded as sufficient evidence for claimed expiry date until data from real time aging studies are available. Fed-Ex Ground, Account number. Shelf Life Study Equivalent. Define package material properties, seal strength and integrity tests, sample sizes, asfm acceptance criteria.

ASTM-F Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices The introduction of new or modified products to the marketplace requires the assurance that they can be stored for an extended period one year, two years, etc.

UPS Ground, Account number. Accelerated Aging Test Results. The sterile barrier system material and device interaction compatibility that may be required for new product development or the resulting evaluation is not addressed in this guide. The sterile barrier system shall maintain sterility to the point of use or until the expiry date. Please ship all samples with this form to: Relative Humidity RH is not 70 factor in the Arrhenius equation.

  IOANICHIE BALAN PDF

Online Quote System – Accelerated Aging. Please fill-in your Input Temperatures and select the Q10 value.

ASTM-F1980

Extracted information from ASTM F testing may be used to support expiration date asmt for medical device sterile barrier systems.

Ambient Temperature Warehouse Temperature Higher temp. UPS 2 nd day, Account number.

The introduction of new or modified products to the marketplace requires the assurance that they can be stored for an extended period one year, two years, etc. 077 Version s – view previous versions of standard. Pricing includes monthly temperature graph reports.

Online Quote System – Accelerated Aging

Accelerated Aging Temperature Higher temp. Stability testing shall demonstrate that the sterile barrier system maintains integrity over time. Link to Active This link will always route to the current Active version of the standard.

Fed-Ex Overnight, Account number.

Therefore, some add an extra month to the aging study to asgm for sterilization, etc. If a different Temperature and or Temperature Humidity condition is required than listed above or more then 30 cubic feet of boxes, contact RMTS for Pricing at Real time studies must be carried out to the claimed shelf life of the product and be performed to their completion.

Determining AAFs are beyond the scope of this guide.